Power Requirements:
Output, RF (Bio Med Only):
Output, RF
(Treatment Mode)
33
Proprietary & Confidential - Page 36 of 391 - Uncontrolled if Printed
Supply Type: Single phase AC.
Protective Electrical Ground
AC Voltage Range: 100-120 / 220-240 V
AC Frequency:
Power: 450 W (max)
Inlet Fuses: T 6.3 A H 250 to IEC 127
Output Frequency: 396.7 kHz +/-2.0 kHz
Power Capability: 200 W
Rated Load: 400 Ohm
Duty Cycle: 10 s ON, 30 s OFF, 1 hour
RF Crest Factor: 1.6 continuous to 3.7 pulsed
Maximum Voltage: 460 V peak
Output Frequency: 396.7 kHz +/-2.0 kHz
Power Capability: 35 W (Max)
Rated Load: 400 Ohm
Duty Cycle: 10 s ON, 30 s OFF, 1 hour
RF Crest Factor: 1.6 continuous to 3.7 pulsed
Maximum Voltage: 460 V peak
Equipment is Class I.
A protective electrical ground
connection must be provided.
50-60 Hz