Specifications
Product Specification Table
Classification Under IEC 60601-1 Concerning Electrical Protection
Equipotential Terminal: According to IEC 60601-1 clause 8.6.7
Applied Part -
Classification to IEC 60601:2005: Type CF. This is Defibrillator Proof.
Environment for
Operation
Environment for
Hospital/Clinic
Storage:
Environment for
Transportation:
Power Supply Cord:
Proprietary & Confidential - Page 35 of 391 - Uncontrolled if Printed
Classification: Class 1. A protective earth connection must be provided
Ambient Temperature:
Relative Humidity: 20 % to 90 % non-condensing
Atmospheric pressure:
Ambient Temperature: +0 °C to +30 °C (+32 °F to +86 °F)
Relative Humidity:
Atmospheric pressure: 69 kPa to 106 kPa
Ambient Temperature: -20°C to +50°C (-4 °F to +122 °F)
Relative Humidity: 10 % to 90 % non-condensing
Atmospheric Pressure: 69 kPa to 106 kPa
-
+10 °C to +30 °C (+50 °F to +86
°F)
≤ 2,000 m (6,560 feet), 80 kPa to
106 kPa
*10 % to 90 % non-condensing
*20 % once unpacked
250 V, 10 A, 2 m long with IEC
60320-C13 type connector.
USA type UL and CSA Recognized