62 W nominal at maximum
Power Capability:
setting measured at the output
socket
Output Frequency: 5,800 MHz +/- 1 MHz
Rated Load: 50 Ohm
RS2 – 10 s ON, 1 s OFF x 3 then
268 s OFF (1-hour max operation
time)
Duty Cycle: (Please refer to
applicable IFU):
HS1 – 10 s ON, 1 s OFF x 5 then
246 s OFF (1-hour max operation
Output, Microwave
time)
(Treatment Mode):
(The duration of the "ON" period
SB1-L
– 5 s ON, 1 s OFF x 5 then
is automatically limited by the
268 s OFF (1-hour max operation
Generator and specific accessory,
time)
re-application of the blue
AB1 – 180 s ON, 30 s OFF x 9
Footswitch pedal is required to
then 28 min 30 sec OFF (1-hour
resume energy activation)
max operation time).
NP1 – 120 s ON, 120 s OFF x 8
then 28 min OFF (1-hour max
operation time).
1.4 (output is a sine wave)
RF Crest Factor:
The software version is displayed
on the Electrosurgical Generator
Software Version:
following switch on from
Standby
Degree of protection:
IP00
Weight: 16 kg / 35 lb
Height 165 mm +/- 5 mm
Physical:
Dimensions:
Width 400 mm +/- 5 mm
Depth 458 mm +/- 5 mm
34
Proprietary & Confidential - Page 37 of 391 - Uncontrolled if Printed