2.6 Installation procedures
2.6.1 Open the package, check whether the equipment is complete as per the packing list,
and place the main unit on a solid place, facing the operator.
2.6.2 8 Plug the external air pipe connector into the air intake connector on the back of the
main unit (as showed in Figure 8).
2.6.3 Fill the water bottle with an appropriate amount of water, and fix the water bottle into
the sink seat directly above the main unit (it is recommended to smear a layer of petro-
leum jelly on the O-ring at the bottom of the water bottle to facilitate the plugging and
unplugging of the water bottle)
2.6.4 Installation of foot pedal
1) Wired mode: Connect one end to the foot pedal port of the foot pedal, and the other
end to the foot pedal port on the main unit.
2) Wireless mode: Remove the pedal battery cover and install 2 AA batteries. Switch
on the main unit and the foot pedal can be connected and controlled (when leaving the
factory, the Bluetooth foot pedal and the main unit have been paired and connected. If
you change the foot pedal, please enter the "System Settings" to match the Bluetooth
foot pedal.)
2.6.5 Connect the ultrasonic handpiece and the air polishing handpiece into the correspond-
ing cables, and place the handpiece on the brackets on both sides of the main unit. The
ultrasonic handpiece is on the left and the air polishing one is on the right.
2.6.6 Turn off the power switch, connect the power cord to the power socket on the back of
the main unit, and then plug in the power, as shown in Figure 2.
Warning 1: When the power cord is connected to the network power supply, it must be
connected to a protective ground; When the power cord is connected to the network
power supply, do not place or install the product where it is difficult to disconnect the
network power supply.
3 Function and operation
3.1 Description of the foot pedal function
Different machines are equipped with different kinds of foot pedal, please refer to the actual
packing list.
3.1.1 Multifunctional foot pedal
a) Choose the mode of foot pedal according to 2.6.4, place it in the flat ground
b) The pedal and its functions are shown below:
Button
Operating mode
Ultrasonic system
A
Standard
Vibration + Water
B
No-water mode
Vibration
C(A+)
Boost mode
5-Level increase in Power
D
Cleaning mode
Water only
Note: Under Boost mode, Power or air pressure would be increased by 5 grades, the maxi-
mum is the 12th grade.
After releasing button C, it will automatically return to the previously set grade.
3.1.2 Standard foot pedal (round)
See description of foot pedal key A in the table above.
3.2 Ultrasonic system
3.2.1 Scaling
1) Turn on the power switch and pick up the ultrasonic handpiece, afterwards the panel
will automatically shift to Ultrasonic mode.
2) Press the G button to enter Scaling mode.
3) Select the tip and tighten it on the handpiece by torque wrench.
4) When stepping on the button A, tip vibrates with LED on and water emits; Release
the button, vibration and water stop, LED will be off 10 seconds later.
5) The handpiece is generally held in a pen-holding position.
6) The frequency is extremely high when the machine is on working. It can remove the
tartar with the tips softly stroke on the teeth surface like erasing motion. DO NOT stay
long or put overmuch pressure on the teeth.
7) Vibrating intensity: Adjust the vibrating intensity according to your need. The setting
power is level 3, please adjust the vibrating intensity according to the teeth sensitivity
and hardness of calculus during clinical application.
8) Water volume: Please adjust the volume on the panel.
9) During clinical scaling, please keep the side of the tip adapt to the tooth surface
horizontally, with light lateral pressure to allow the tip vibrate freely.
10) After operation, please keep device working for 30s with water to flush the hand-
_R0_XPOWER ablatore_bug_6lingue.indd 5
_R0_XPOWER ablatore_bug_6lingue.indd 5
Function
Air polishing system
AirPowder + water
Air only
5-Level increase in air pressure
Water + Air
Figure 9.
Schematic diagram of multi-function foot pedal
Figure 10.
Schematic diagram of multi-function foot pedal
piece and tips; Remove the tip and get it sterilized.
3.2.2 Periodontal treatment
1) Use a torque wrench to tighten the periodontal treatment tip to the ultrasonic hand-
piece. Click the P button on panel to enter the periodontal treatment mode.
2) The operation and adjustment methods are similar to Scaling mode.
3.2.3 Endodontic irrigation
1) Tighten the endo file to the handpiece by endo wrench.
2) Click on the E button and enter the Endo mode.
3) The default power is level 1 under Endo mode, operator can adjust the power ac-
cording to the need in clinical treatment.
4) Select the appropriate Endo file and slowly place it into the root canal of the patients
teeth. Step on the foot pedal to start ultrasonic endodontic irrigation.
5) During clinical cleansing, the file should not be pressed too tightly when it is in the
root canal
6) Do not step on the pedal before the file is in the root canal.
7) The suggested power range for endodontic irrigation is level 1-level 5.
3.3 Air polishing system
Use exclusively Xpower powders.
FAST type powders (65µm) must be loaded into FAST pwder tank only and used exclusively
with PA handpiece.
MILD tupe powders (25/14µm) must be loaded into MILD pwder tank only and can be used
with both PA and PAP handpieces.
1) Add appropriate amount of powder to the powder tank (the amount of powder added
should be between "Max" and "Min" marked on the the tank), then tighten the upper cover
of the powder tank, and fix the powder tank on the main unit.
2) Pick up the Air polishing handpiece, afterwards the panel will automatically shift to Air
polishing mode. Select powder type (FAST/MILD) on display.
3) Adjust the water volume and air pressure, aim the nozzle at a pool, press the foot button A,
and confirm that the nozzle can spray gas, powder and water spray normally before use.
4) Before Air polishing treatment, please wear goggles and veil on the face of the patient.
Users should wear goggles or a protective mask.
5) The handpiece is generally held in a pen-holding position.
6) Adjust the water volume and air pressure to an appropriate level. The recommended water
volume starts at level 5 and the air pressure starts at level 3. According to the sensitivity
of the patient's teeth and hardness of the dental plaque, adjust the water volume and air
pressure at any time during the clinical process; increasing the air pressure will enhance
cleaning effect, but will weaken the polishing effect; increasing the amount of water will
enhance the polishing effect, but will weaken the cleaning effect.
7) The nozzle should be aligned with the tooth surface during the cleansing, but avoid direct
contact. Keep the nozzle and the tooth surface at a distance of 3-5mm at 30-60 angle.
The smaller the angle, the larger the cleaning area; during the cleansing, please Perform
a small circular motion on the surface of the tooth.
8) The air/powder mixture reflected from the tooth surface should be evacuated by using a
strong suction device on the dental unit during treatment.
9) After treatment, adjust the water volume to the maximum level, and polish the surface of
teeth.
3.4 Cleaning mode
To keep the unit pipeline clean and fresh it is recommended to use Clean&White Water Jet
Tablets into water tank.
It is recommended to flush and disinfect the pipeline of the device every day. "Cleaning"
mode allows cleaning and disinfection of the pipeline to reduce the accumulation of crystals
and the number of bacteria in the pipeline.
1) Fill a water tank with distilled water or demineralized water.
2) Pick up the ultrasonic handpiece, point the handpiece at the sink, click the Cleaning button
on the screen, Step on the pedal to begin cleaning.
3) 30s after cleaning, release the pedal to stop.
4) After cleaning, put the ultrasonic handpiece back into the bracket. And then, pick up the air
polishing handpiece, point the handpiece nozzle at the pool, and click the Cleaning button
again, so that the device will automatically blow out the residual powder in the pipeline and
release the high pressure gas in powder tank.
5) After cleaning for 15 seconds, the device will automatically exit the Cleaning mode. You
can also click Purge on the screen to stop cleaning.
3.5 Setting
1) Click "ON" /"OFF" to turn on or off the heating function. (According to the actual water
output, the water temperature is different).
2) Click the "Settings" button at the bottom right corner of the screen to enter the "System
Settings" interface
3) Click Calibration button, the system will automatically calibrate the air pressure to the
value set in the program
4) The Bluetooth pedal has been matched to the machine, any subsequent replacement of
foot pedal or main unit needs to make new pair.
EN
| 5
Instruction Manual