See "Appendix A: Examples of MRI-CS eligibility sheets" on page 24.
Yes
(1) Confirm the patient's name and date on the eligibility sheet. The date on
the eligibility sheet should be on or near the date of the MRI
appointment.
Note: The further the date on the eligibility sheet is from the patient's MRI
appointment, the greater the chance that the following occurred:
(2) Go to step 4 on page 10.
No
Go to the next step.
2.
Model number of patient control device
Did the patient bring a Model 977__ series patient control device to the MRI
appointment?
Yes
(1) Ask the patient to navigate to the MRI Mode screen to activate MRI
mode.
(2) After the MRI Mode screen displays, go to step 4 on page 10.
No
Go to the next step.
3. Does the patient have a Medtronic patient ID card with a neurostimulator model number
that matches one of the neurostimulator model numbers listed on page 5?
Yes
Go to "Head-only eligible MRI scan conditions" on page 18.
The patient had an event (eg, revision surgery of the implanted
neurostimulation system) that may have changed the scan eligibility.
The patient's stimulation was turned back on.
MRI mode: When MRI mode is activated:
– The screen on the patient control device displays MRI-CS scan-
type eligibility.
– If the eligibility screen displays, it means stimulation has been
turned off.
If the patient is unable to navigate to the MRI Mode screen, contact
Medtronic Technical Services for assistance.
Do not deactivate, or exit, MRI mode or turn stimulation on with the
patient control device until after the patient's MRI scan is complete and
the patient is outside of the scanner (magnet) room.