Auto  pressure  zeroing  
Cuff  auto  deflation  
Inflation  time  for  cuff  
Measurement  time  
Initial  inflation  pressure  
Software  overpressure  
protection  
Intervals  for  AUTO  
measurement  time  
Alarm  range  
Alarm  Indication  
Measure  mode  
PR  
PR  range  
Resolution  
Accuracy    
Recovery  time  after  
defibrillation  
 
 
70
The  device  will  automatically  zero  itself  as  soon  as  it  is  turned  on.  
The  cuff  will  deflate  automatically  when  power  is  off  or  the  time  of  measurement  
exceeds  120  seconds  (90  seconds  for  neonate)  or  the  cuff  pressure  exceeds  the  
overpressure  protection  levels,  set  by  software  and  hardware.  
<40s  (standard  adult  cuff)  
Normally,  it  is  20s~45s  (depending  on  HR  and  interference  from  movement)  
Adult  default:  160  mmHg  
Pediatric  default:  130  mmHg  
Neonatal  default:  75  mmHg  
Double  hardware,  software  overpressure  protection  
Adult:  (297±3)  mmHg  
Pediatric:  (252±  3)  mmHg  
Neonatal:  (147±3)  mmHg  
5min-­240min  
0  mmHg  to  300  mmHg,  high/low  limit  can  be  adjusted  
Sys  
continuously.  
0  mmHg  to  300  mmHg,  high/low  limit  can  be  adjusted  
Dia  
continuously.  
0  mmHg  to  300  mmHg,  high/low  limit  can  be  adjusted  
Map  
continuously.  
Three  levels  of  alarms:  sound-­light  alarms,  color  changes  in  alarm  limits  area;;  and  
alarms  with  text  prompts.  
Adult  
Single,  Cycle,  STAT,  Average  
Pediatric  
Single,  Cycle,  STAT,  Average  
Neonatal:  
Single,  Cycle,  Average  
40  bpm  to  240  bpm  
1  bpm  
±3  bpm  
<5s  
 
28