Auto pressure zeroing
Cuff auto deflation
Inflation time for cuff
Measurement time
Initial inflation pressure
Software overpressure
protection
Intervals for AUTO
measurement time
Alarm range
Alarm Indication
Measure mode
PR
PR range
Resolution
Accuracy
Recovery time after
defibrillation
70
The device will automatically zero itself as soon as it is turned on.
The cuff will deflate automatically when power is off or the time of measurement
exceeds 120 seconds (90 seconds for neonate) or the cuff pressure exceeds the
overpressure protection levels, set by software and hardware.
<40s (standard adult cuff)
Normally, it is 20s~45s (depending on HR and interference from movement)
Adult default: 160 mmHg
Pediatric default: 130 mmHg
Neonatal default: 75 mmHg
Double hardware, software overpressure protection
Adult: (297±3) mmHg
Pediatric: (252± 3) mmHg
Neonatal: (147±3) mmHg
5min-240min
0 mmHg to 300 mmHg, high/low limit can be adjusted
Sys
continuously.
0 mmHg to 300 mmHg, high/low limit can be adjusted
Dia
continuously.
0 mmHg to 300 mmHg, high/low limit can be adjusted
Map
continuously.
Three levels of alarms: sound-light alarms, color changes in alarm limits area;; and
alarms with text prompts.
Adult
Single, Cycle, STAT, Average
Pediatric
Single, Cycle, STAT, Average
Neonatal:
Single, Cycle, Average
40 bpm to 240 bpm
1 bpm
±3 bpm
<5s
28