for  5  cycles  
70%  UT  
(30%  dip  in  UT)  
for  25  cycles  
<5%  UT    
(>95%  dip  in  UT)  
for  5  sec  
Power  frequency  (50Hz)  
3  A/m  
magnetic  field  
IEC  61000-­4-­8  
NOTE:      UT  is  the  A.C.  mains  voltage  prior  to  application  of  the  test  level.  
Guidance  and  manufacturer's  declaration  –  electromagnetic  immunity  
Guidance  and  manufacturer's  declaration  –  electromagnetic  immunity  
–for  EQUIPMENT  and  SYSTEMS  that  are  not  LIFE-­SUPPORTING  
Guidance  and  manufacture's  declaration  –  electromagnetic  immunity  
The  monitor  is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The  customer  or  the  user  of  monitor  should  
assure  that  it  is  used  in  such  an  environment.  
Immunity  test  
IEC  60601  test  level  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Conducted  RF  
3  Vrms  
IEC  61000-­4-­6  
150  kHz  to  80  MHz  
 
 
Radiated  RF  
3  V/m  
IEC  61000-­4-­3  
80  MHz  to  2.5  GHz  
 
 
 
NOTE  1  
At  80  MHz  and  800  MHz,  the  higher  frequency  range  applies.  
NOTE  2  
These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is  affected  by  absorption  and  
reflection  from  structures,  objects  and  people.  
a.  Field  strengths  from  fixed  transmitters,  such  as  base  stations  for  radio  (cellular/cordless)  telephones  and  land  mobile  radios,  
amateur  radio,  AM  and  FM  radio  broadcast  and  TV  broadcast  cannot  be  predicted  theoretically  with  accuracy.  To  assess  the  
electromagnetic  environment  due  to  fixed  RF  transmitters,  an  electromagnetic  site  survey  should  be  considered.  If  the  measured  
field  strength  in  the  location  in  which  the  monitor  is  used  exceeds  the  applicable  RF  compliance  level  above,  the  monitor  should  
be  observed  to  verify  normal  operation.  If  abnormal  performance  is  observed,  additional  measures  may  be  necessary,  such  as  
reorienting  or  relocating  the  monitor  
 b.  
Over  the  frequency  range  150  kHz  to  80  MHz,  field  strengths  should  be  less  than  3  V/m.  
Recommended  separation  distances  between  portable  and  mobile  RF  communications  equipment  and  the  
EQUIPMENT  or  SYSTEM  –  for  EQUIPMENT  or  SYSTEM  that  are  not  LIFE-­SUPPORTING  
portable  and  mobile  RF  communications  equipment  and  the  Q3  monitor  
The  monitor  is  intended  for  use  in  an  electromagnetic  environment  in  which  radiated  RF  disturbances  are  controlled.  The  
for  5  cycles  
70%  UT  
(30%  dip  in  UT)  
for  25  cycles  
<5%  UT    
(>95%  dip  in  UT)  
for  5  sec  
3  A/m  
–for  all  EQUIPMENT  and  SYSTEMS  
 
Compliance  level  
 
 
 
 
 
 
 
3  Vrms  
 
 
3  V/m  
34
 
Recommended  separation  distances  between  
Power  frequency  magnetic  fields  
should  be  at  levels  characteristic  of  a  
typical  location  in  a  typical  commercial  
or  hospital  environment.  
Electromagnetic  environment  -­  guidance  
Portable  and  mobile  RF  communications  equipment  
should  be  used  no  closer  to  any  part  of  the  monitor  
including  cables,  than  the  recommended  separation  
distance  calculated  from  the  equation  applicable  to  
the  frequency  of  the  transmitter.  
Recommended  separation  distance  
 
Where  P  is  the  maximum  output  power  rating  of  the  
transmitter  in  watts  (W)  according  to  the  transmitter  
manufacturer  and  d  is  the  recommended  separation  
distance  in  metres  (m).  
Field  strengths  from  fixed  RF  transmitters,  as  
determined  by  an  electromagnetic  site  survey,  
should  be  less  than  the  compliance  level  in  each  
frequency  range.  
Interference  may  occur  in  the  vicinity  of  equipment  
marked  with  the  following  symbol:
 
 
 
75