Irregularity
Description
Power fails
to come on
No video image
aBox™ Duodeno
does not boot up
Power fails
to come on
aBox™ Duodeno will
not shut down
Rinsing not possible
Possible Cause
The power switch and/or the
power button of the aBox™
Duodeno is set to OFF
The power cable
is not connected
The aBox™ Duodeno
or ancillary equipment
is not switched ON
Medical grade monitor
not properly connected
or defective
aScope™ Duodeno
not properly connected
or defective
LEDs not switched ON
aBox™ Duodeno
not powered ON
Medical grade monitor
not properly connected
or defective
Power cable not connected
Fuse blown
aScope™ Duodeno
still connected
Power button was pressed
for too short a duration
aBox™ Duodeno defective
Peristaltic pump defective
Solution
Set the power switch to ON
Connect the power cable with an
electrical source as described in
section 5
Switch aBox™ Duodeno and
ancillary equipment on
Connect the medical grade
monitor properly
Connect the aScope™ Duodeno
correctly or connect a new
aScope™ Duodeno
Switch the LEDs ON
Power the aBox™ Duodeno ON
Shutdown aBox™ Duodeno
and connect the medical grade
monitor properly. Try to power
ON the aBox™ Duodeno after
10 sec again
Connect the power cable
Replace fuse. Please refer
to section 7
Please unplug the aScope™
Duodeno from the aBox™ Duodeno
and press the power button for
3 secs. to shut down the system
Press the power button
for at least 3 sec
Switch the aBox™ Duodeno off by
using the mains switch on the
back panel of the aBox™ Duodeno
and contact Ambu for support
Replace the peristaltic pump.
See section 8