Guidelines and manufacturer's declaration – resistance to electromagnetic interference.
The aScope™ Duodeno is intended for use in the electromagnetic environment specified
below. The user of the aScope™ Duodeno should ensure that it is used in such an environment.
Transmission measurement
HF transmission according
to CISPR 11
HF transmission according
to CISPR 11
Upper harmonics according
to IEC 61000 – 2 – 3
Voltage fluctuations / flicker
according to IEC 61000 – 3 – 3
13. Contact
Manufacturer
Ambu A/S
Baltorpbakken 13
2750 Ballerup, Denmark
Appendix 1. Connection Configurations Descriptions
The aBox™ Duodeno is equipped with multiple video outputs for viewing by clinicians and
several options for documentation via additional video outputs and a remote signal (i.e.
"trigger") to activate common recording devices. The steps in this appendix will guide the
installer through various available connections and the required cabling for each configuration.
Table 4 Electromagnetic Transmission
Compliance
Group 1
Class A
Class A
Satisfied
Electromagnetic
Environmental Guidelines.
The aScope™ Duodeno uses HF
power internally only. Its HF
transmission is therefore very low
and it is improbable that it will
cause interference with electrical
equipment in its vicinity.
The aScope™ Duodeno is suitable
for use in facilities other than
residential areas that are
connected directly to the public
power grid that also supplies
buildings used for residential
purpose, provided that the
following warning is observed:
Warning:
This device is intended only for use
by trained medical personnel. This
is a Class A device according to
CISPR 11. In a residential area, this
unit can cause radio interference,
so it is necessary in this case to
take suitable corrective measures,
such as readjusting it, rearranging
it, or shielding the unit or filtering
its power connection.
25