Technical Data - Ambu aBox Duodeno Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 161
8. Peristaltic Pump Replacement
Ensure the aBox™ Duodeno is powered off and that the peristaltic pump housing is
completely closed.
Release the entire peristaltic pump head by pressing the release tab on the lower-right
side of the pump housing. 1
Rotate the peristaltic pump to your left and pull it towards you in one continuous
motion. 2
Align the key of the replacement peristaltic pump assembly with the corresponding
notch on the aBox™ Duodeno, then push and rotate the pump assembly to the right until
it locks into position. 3
When properly secured, peristaltic pump head should not rotate in any direction. 4
1
9. Cleaning of the aBox™ Duodeno External Surfaces
The aBox™ Duodeno should be cleaned before and after each procedure. Clean the aBox™
Duodeno according to good medical practice and using the procedures below:
Use a germicidal wipe to remove heavy soil. All blood and other body fluids must be
thoroughly cleaned from all surfaces and objects before using germicidal wipe.
(SUPER- SANI-CLOTH® from PDI or equivalent)
The treated surfaces must remain visibly wet for the full length of two (2) minutes. Use
additional wipes, if needed, to assure continuous 2 minutes wet contact time. Let the aBox™
Duodeno air dry.

10. Technical Data

All the following reported measurements (e.g. weight, dimensions) are average values.
Therefore, small variations may occur, which however have no effect on the performance and
safety of the system.
10.1. aBox™ Duodeno Specifications
Power Supply
Size of the aBox™ Duodeno
2
Voltage
Frequency
Consumption
electric power
Fuse rating
Fuse size
Dimensions
Weight
3
AC 120 V / 230 V
50 / 60 Hz
91 VA
2x 5 A H 250 V T
5 mm x 20 mm
494 (D) x 487 (W) x 145 (H) mm
13 kg
4
9
17

Publicidad

Tabla de contenido
loading

Tabla de contenido