infection, it is strictly forbidden to ship contaminated medical devices. The medical device
aScope™ Duodeno must be decontaminated on site before shipment to Ambu. Ambu reserve
the right to return contaminated medical devices to the sender.
In an event of a serious incident, please inform Ambu and the competent authority.
Disposal of the aScope™ Duodeno
The Single-Use aScope™ Duodeno is considered medical waste after use and must be disposed
of in accordance with local guidelines for medical waste with electronic components. This is
a single-use device, so do not soak, rinse or sterilize as it may leave harmful residues or cause
malfunction of the device. The design and material used are not compatible with conventional
cleaning and sterilization procedures.
7. Technical Data
All the following reported measurements (e.g. weight, dimensions) are average values.
Therefore, small variations may occur, which however will have no effect on the performance
and/or safety of the system.
7.1. aScope™ Duodeno
Endoscope Functions
Total Length
Insertion section
Bending section
Illumination method
Optical System
Working channel inner diameter
7.2. Storage and Operation Environment Specifications
Transportation temperature
Storage temperature
Operation temperature
Relative humidity for transportation,
storage and operation
Atmospheric pressure
3100 mm
Working length
Insertion tube outer diameter
Distal end outer diameter
Angulation
LED
Field of view
Direction of view
Ø 12.6 Fr (4.2 mm)
-5 °C – +40 °C (23 °F – 104 °F)
10 °C - 25 °C (50 °F - 77 °F)
10 °C - 40 °C (50 °F - 104°F)
30 % – 85 % relative
80 kPa – 109 kPa (100 kPa=1 Bar)
11.6 – 15.8 psi
600 – 818 mmHg
1240 mm
11.3 mm
13.7 mm
120° Up
90° Down
90° Left
110° Right
130°
Backward side
viewing 6°
19