5. When the program is finished, remove the Deflector Shaft from the washer-
disinfector unit.
6. After cleaning and disinfection, open the lid of the cleaning and sterilization
tray or remove the Deflector Shaft Cleaning Clip from the Abrasor/Resector
locking button of the product and disconnect the rinsing tubing of the washer-
disinfector unit from the Luer-Lock connection of the Deflector Shaft.
7. Take the Deflector Shaft out of the tray and inspect it according to chapter
"12.1 Control of Function".
8. After the functional testing, put the Deflector Shaft in a suitable sterile bar-
rier system for sterilization (see chapter "13 Packaging"). Apply an additional
drying time in a clean place, if necessary.
joimax
recommendation regarding cleaning and disinfection: Proof of
®
fundamental product suitability for effective mechanical cleaning and disin-
fection was provided by an independent accredited testing laboratory with the
use of the washer-disinfector unit WD BHT INNOVA
BHT Hygienetechnik GmbH, Gersthofen) and the cleaning agent Neo-disher me-
diclean forte (Dr. Weigert GmbH & Co. KG, Hamburg), whereby the procedure
described above was used.
11.4 MANUAL CLEANING AND DISINFECTION
The manufacturer has not validated a manual cleaning and disinfection pro-
cedure for the Deflector Shaft. In case the customer wants to use a manual
cleaning and disinfection procedure for the Deflector Shaft, the responsibility
for achieving the desired result by performing a respective manual cleaning and
disinfection procedure is with the respective processor. A verification and/or
validation and routine monitoring of the manual procedure is required.
12. CONTROL OF FUNCTION AND MAINTENANCE
12.1 CONTROL OF FUNCTION
After cleaning/disinfection, check the Deflector Shaft for function, corrosion,
damaged surfaces, chipping or/and cleanliness.
Furthermore, perform a functional test. The following points need to be taken
care of:
• The bending mechanism of the tip of the Deflector Shaft - using the deflection
adjusting lever – must work accurately. The tip must be deflectable starting
in straight position (0°) in three steps into 12°, 24° and 36° position (watch
reference mark – see figure. 1 or 9 ).
• The Abrasor/Resector-Locking button can be pressed easily and returns auto-
matically after release into its initial position.
• The distal coupling part must be smoothly rotatable.
M3 (thermal disinfection,
®
19