Descargar Imprimir esta página

Boston Scientific MAESTRO 4000 Manual De Instrucciones página 15

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 38
Compatible Catheters
• WARNING: Refer to the Directions for Use (DFU) provided with individual catheters to determine if the
catheters are compatible with the Maestro 4000™ Cardiac Ablation System. Carefully review the specific
indications, contraindications, warnings, precautions and adverse events included in a catheter's DFU
prior to using the catheter with the Maestro 4000 Cardiac Ablation System.
Catheters tested for compatibility for use with the Maestro 4000 Cardiac Ablation System include the BSC
Models listed below.
Note: some catheters may not be available in all geographic areas.
Standard Catheters:
• Blazer II™
• Blazer II HTD™
• Blazer Prime™ HTD
High-Power Catheters:
• Blazer II XP™
• Blazer Prime™ XP
• IntellaTip MiFi XP
Closed-Irrigation Catheters:
• Chilli II Cooled Ablation Catheter
Open-Irrigated Catheters:
• Blazer Open-Irrigated Ablation Catheter
The RF Generator recognizes the catheter/pod combination and automatically adjusts the ranges of power and
temperature settings available for the selected control mode.
Catheter Type
Pod Type
Standard
Standard
High-Power
High-Power
Closed-Irrigation
Closed-Irrigation
Open-Irrigated
Open-Irrigated
Maestro 4000™ Cardiac Ablation System Operator's Manual, Global
Table 1 Adjustment Ranges for Power and Temperature
Temperature Range
Temp-Control Mode
100W
30 – 90 °C
150W
30 – 90 °C
100W
30 – 80 °C
150W
30 – 90 °C
100W
30 – 90 °C
150W
30 – 90 °C
100W
30 – 50 °C
150W
30 – 50 °C
Temperature Range
Power-Control Mode
30 – 95 °C
30 – 95 °C
30 – 85 °C
30 – 95 °C
30 – 95 °C
30 – 95 °C
30 – 50 °C
30 – 50 °C
Power Range
0 - 50 W
0 - 100 W
0 - 100 W
0 - 150 W
0 - 50 W
0 - 50 W
0 - 50 W
0 - 50 W
15
Black (K) ∆E ≤5.0

Publicidad

loading