Descargar Imprimir esta página

Boston Scientific MAESTRO 4000 Manual De Instrucciones página 34

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 38
Code
Display
D05
EXCESSIVE
TEMPERATURE
D06
TEMP OUT
OF RANGE
[15C-95C]
D07
TEMPERATURE
TOO HIGH
D08
CHECK POD
D09
N/A
34
Black (K) ∆E ≤5.0
Description
Measured temperature
greater than 95 °C for more
than one second or measured
temperature exceeds setting in
Power Control mode.
Measured temperature
<15 °C or >95 °C
An attempt to RF delivery with a
measured temperature > 43 °C
or greater than the temperature
setting within 30 seconds of the
previous RF delivery.
Generator detects a pod failure
or an issue with the generator-
to-pod connection.
Invalid Code
Action
Check temperature and power settings.
Check control-mode setting.
Check cables and connectors in the RF pathway
for possible damage or loose connection.
Verify that temperature in standby is
appropriate and doesn't change with agitating
connectors/cables in the RF pathway. If
temperature changes with agitation, replace
suspected components one at a time.
If error recurs, replace catheter cable.
If error recurs, replace Pod.
If error recurs, replace the Generator.
This condition may have been caused by
switching catheters during a procedure.
Check cables and connectors in the RF pathway
for possible damage or loose connection.
Verify that temperature in standby is
appropriate and doesn't change with agitating
connectors/cables in the RF pathway. If
temperature changes with agitation, replace
suspected components one at a time.
If error recurs, replace catheter cable.
If error recurs, replace catheter.
If error recurs, replace Pod.
If error recurs, replace the Generator.
Check temperature setting.
Check cables and connectors in the RF pathway
for possible damage or loose connection.
Verify that temperature in standby is
appropriate and doesn't change with agitating
connectors/cables in the RF pathway. If
temperature changes with agitation, replace
suspected components one at a time.
If error recurs, replace catheter cable.
If error recurs, replace Pod.
If error recurs, replace the Generator.
Check generator-to-pod connection.
If error recurs, replace pod.
Contact BSC
Maestro 4000™ Cardiac Ablation System Operator's Manual, Global

Publicidad

loading