Descargar Imprimir esta página

Service And Maintenance - Boston Scientific MAESTRO 4000 Manual De Instrucciones

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 38
RF Power Delivery with all other Catheters
Starting RF Power Delivery:
• Press the RF Power Control button once, or hold down the Foot Switch pedal.
• Delivery of RF Power will start immediately.
• The RF Power Control button will illuminate and remain lit until RF power delivery stops.
• RF Power will stop automatically when the setting for maximum duration of RF Power Delivery is
reached.
Stopping RF Power Delivery:
• To stop delivery of RF Power before reaching the setting for maximum duration, press the RF
Power Control Button, or release the Foot Switch pedal.
• The RF Power Control button will flash to indicate that the RF Generator is in STANDBY state.

SERVICE AND MAINTENANCE

Preventative Inspection
• WARNING: Damage such as frayed cords or cables and cracks or dents on the equipment may result in
electrical shock.
• Caution: BSC recommends that the RF Generator and Remote be powered off at the end of each
procedure in order to ensure that the self-test is performed before the next procedure.
During the useful life of the equipment, maintain close watch for damage such as frayed cords or cables
and cracks or dents on the equipment. If damage is identified, take the equipment out of service and contact
Boston Scientific Corporation for service requirements.
Functional Self-Test
System does not require periodic preventative maintenance and calibration. Upon start up, the unit performs
self-test per OPERATION Start-up Section of this manual. For further details refer to that section.
System Servicing
None of the System components are user-serviceable. Contact Boston Scientific Corporation for all service
requirements.
Cleaning/Disinfecting
• WARNING: Do not immerse the Generator, the Remote, or accessories in any liquid.
• WARNING: Use of non-flammable cleaning and disinfection agents is recommended. If used, flammable
agents or solvents should be allowed to evaporate before high-frequency surgery. There is a risk of
pooling of flammable solutions under the PATIENT or in body depressions such as the umbilicus, and
in body cavities such as the vagina. Any fluid pooled in these areas should be mopped up before HF
SURGICAL EQUIPMENT is used.
• Caution: The Generator, Pod, Remote, Footswitch, power cords, and communication cables are not
intended to be sterilized and should remain outside of the sterile field.
The outer surfaces of the Generator, the Pod, the Remote and their accessories may be wiped clean with a mild
soapy solution. Isopropyl alcohol may be used to clean the outer surfaces. Avoid caustic or abrasive cleaners.
20
Black (K) ∆E ≤5.0
Maestro 4000™ Cardiac Ablation System Operator's Manual, Global

Publicidad

loading