Descargar Imprimir esta página

Boston Scientific MAESTRO 4000 Manual De Instrucciones página 9

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 38
Average button
• When the Average Button is pressed in STANDBY state, the average values of power, temperature
and impedance for the previous ablation are shown in the alphanumeric display.
• When the Average Button is pressed in DELIVER state, the average values of power, temperature and
impedance for the current ablation are shown in the alphanumeric display.
Maximum button
• When the Maximum Button is pressed in STANDBY state, the maximum values of power, temperature
and impedance for the previous ablation are shown in the alphanumeric display.
• When the Maximum Button is pressed in DELIVER state, the maximum values of power, temperature
and impedance for the current ablation are displayed in the alphanumeric display.
Isolated Patient Connector
Provides for connection of the Pod to the RF Generator.
Front Panel Indicators
The front panel indicators include the following:
System Fault
Temperature Mode
Power Mode
RF Power Control
RF Power Control
Memory
Panel Active
Panel Inactive
Average
Maximum
Remote (Optional Accessory)
The Remote allows an operator access to all controls, displays, and indicators while at a distance from the
RF Generator. The Remote can be connected to the RF Generator with cables of various lengths. All controls,
displays, and indicators on the Remote have the same appearance and function as those on the RF Generator.
Foot Switch (Optional Accessory)
A Foot Switch with a 10 foot cable can be connected to the RF Generator. The Foot Switch provides hands-free
control to start/stop RF delivery. The Foot Switch functions similarly to the RF Power Control button except that
the operator must continuously hold the Foot Switch down for RF power to be delivered. RF power delivery is
immediately terminated when the operator's foot is lifted off the Foot Switch.
Pressing and releasing the Foot Switch can also be used to clear diagnostic messages and exit diagnostic state.
Maestro 4000™ Cardiac Ablation System Operator's Manual, Global
Red
Indicates a System Fault has occurred.
Green
Indicates System is in Temperature-Control Mode
Green
Indicates System is in Power-Control Mode
Flashing Amber
Indicates System is in STANDBY state
Solid Amber
Indicates System is in RF DELIVER state
Green
Indicates settings are loaded from memory
Green
Indicates the Control Panel is active
Yellow
Indicates the Control Panel is inactive
Green
Indicates the Average feature is active
Green
Indicates the Maximum feature is active
9
Black (K) ∆E ≤5.0

Publicidad

loading