Code
Display
D10
INVALID
CATHETER ID
D11
CHECK
REMOTE
D12
CHECK PUMP
N/A
SYSTEM FAULT An unrecoverable error has
Maestro 4000™ Cardiac Ablation System Operator's Manual, Global
Description
Invalid catheter ID
There is a communications
problem between the Remote
and the Generator or the Remote
was disconnected or connected
during RF.
There is a communication error
with the MetriQ™ Pump or there
is a diagnostic condition at the
Pump.
occurred on the Generator/
Remote (as indicated by which
unit displayed the "SYSTEM
FAULT" LED). The Generator/
Remote enters SAFE mode and
will not allow any operations.
Action
Check catheter cable for damage.
If error recurs, replace catheter cable.
If error recurs, replace catheter.
If error recurs, replace catheter Pod.
Do not disconnect or connect the Remote
during RF.
Check cable and connections between the
Generator and remote, tightening the thumb
screws on the connectors, or replacing
component if damage is suspected.
Clear diagnostic condition at Pump.
Check cable and connections between the
Generator and Pump, tightening the thumb
screws on the connectors, or replacing
component if damage is suspected.
Power off the Generator/Remote and restart.
If error recurs, contact BSC field service for
further assistance.