Parameter
Settings
ST.DEL.
SHORT*
MEDIUM
LONG
NONE
ZERO/TAR.
W/ STAB.*
AT STAB.
W/O STB.
AUTOZER.
ON*
OFF
UNIT
GRAMS
CARATS
GRAINS
PENYWT.
MILLIGR.*
MOMMES
TOLA
MESGHAL
DISP.DIG.
ALL*
LP.ON/OFF
DIVIS. 1
MINUS 1
CAL.-ADJ.
INT.CAL.*
INT.LIN.
EXT.CAL.
E.CAL.USR.
CAL.SEQ.
ADJUST*
CAL.-ADJ.
* Factory setting
Explanation
Sets the stability delay to "short": The stability symbol only appears after a short
delay in order to provide a reliable result despite fluctuations.
Sets the stability delay to "medium": The stability symbol only appears after a
longer delay in order to provide a reliable result in case of higher fluctuations.
Sets the stability delay to "long": The stability symbol only appears after a long
delay in order to balance out major instability.
Sets the stability delay to "none": The stability symbol is displayed after the
stability criterion is reached.
After stability: The function of the [Zero] or [Tare] key is only executed after
stability is achieved.
At stability: The function of the [Zero] or [Tare] key is executed if stability exists
when the key is pressed.
Without stability: The function of the [Zero] or [Tare] key is executed
immediately once the key is pressed.
Activates automatic zeroing. The display is automatically set to zero in case of a
deviation of 0 less than (X).
Deactivates automatic zeroing. Zeroing must be triggered with the [Zero] key.
— The device displays the selected unit.
— The availability of units may depend on national legislation and is therefore
country-specific.
"All digits on": All digits are shown in the display. Not available on conformity-
assessed devices.
"Reduced by 1 decimal place for load change": The last decimal place on the
display is switched off until stability is achieved.
"Last digit scale interval 1": The last digit always shows scale interval 1.
"Last decimal place off": The last digit is switched off.
The [Adjust] button starts an internal calibration.
The [Adjust] button linearizes the balance with the integrated weight (not for
precision balance).
The [Adjust] button starts an external calibration with the preset calibration
weight.
The [Adjust] button starts an external calibration with the user-defined
calibration weight value.
Calibration and adjustment is one routine.
Adjustment must be started or exited manually after calibration with the
[Adjust] button.