8.5.1
Adjusting with the isoCAL Function
The device can be automatically internally calibrated and adjusted using the
isoCAL function.
Requirements
— The isoCAL function is set up (for the configuration values for the
"ISOCAL" parameter, see Chapter "4.8.1 Parameters in the "SETUP/
BALANCE" Menu", page 30).
— The conditions for triggering and executing the isoCAL function are met
(see Chapter "15.9 Conditions for isoCAL Function", page 81).
Procedure
If the isoCAL function is set to automatic start and the isoCAL function is
t
triggered:
The [isoCAL] button flashes in the operating display.
y
Wait until the isoCAL function is executed.
t
In the operating display, a time display counts down from 15 seconds
y
to 0.
If no load change or no operation takes place on the device before
y
the expiration of the time display: The isoCAL function starts.
If the isoCAL function is set to manual start and the isoCAL function is
t
triggered:
The [isoCAL] button flashes in the operating display.
y
Press the [isoCAL] button.
t
The isoCAL function starts.
y
8.5.2
Internally Calibrating and Adjusting the Device
Internal Calibration and Automatic Adjustment
Requirements
— For internal calibration: The configuration value "INT.CAL." is set in the
"SETUP" / "BALANCE" / „CAL/ADJ." menu.
— For the adjustment function based on calibration and automatic
adjustment: The "JUST." configuration value is set in the
"SETUP " / "BALANCE" / "CAL.SEQ." menu.
— The weighing pan is unloaded.
— The displayed weight value remains unchanged.
Procedure
Press the [Zero] key.
t
To start the internal calibration/adjustment: Press the [Adjust] button.
t
Operating Instructions Cubis