8.6
Weighing
When weighing chemicals, suitable vessels must be used for the material to
be weighed. This makes it possible to prevent damage to the device or its
accessories.
Requirements
The device is leveled and adjusted.
Procedure
Press the [Zero] key.
t
If below-balance weighing is being carried out: Suspend the sample on
t
the hook for below-balance weighing, e.g. with a wire.
If a vessel is being used for the sample:
t
Place the vessel for the material to be weighed onto the weighing
t
pan.
Press the [Tare] key. This compensates for the weight of the vessel.
t
Place the sample in the vessel.
t
If no vessel is used for the sample: Place the sample on the weighing pan.
t
If the weight value no longer changes and the weighing unit is displayed:
t
Read off the measured value.
8.7
Printing Results
8.7.1
Printing the Calibration Results
The results of the calibration can be printed in a GLP printout.
The use of the isoCAL function must be set (for configuration values for the
"GLP" parameter, see Chapter "4.8.12 Parameters in the "DATA.OUTP./
PRNT.PARA." Menu", page 38):
Procedure
Calibrate the device.
t
Once calibration is complete:
y
The printing process starts.
y
The GLP footer is printed.
y
Operating Instructions Cubis
Operation
| MCE Models
61
®