Operation
60
Operating Instructions Cubis
The calibration and adjustment function is executed:
y
— If the adjustment weight placed on the device is within the specified
limits: The device is automatically adjusted after the calibration.
— If the weight placed on the device is too high: The "+" symbol is
displayed.
— If the weight placed on the device is too low: The "-" symbol is
displayed.
— The [CAL.END.] message indicates the end of calibration.
Remove the adjustment weight.
t
Starting External Calibration And Adjustment Manually
An adjustment weight must be placed on the weighing pan. The weight
value for the adjustment weight is displayed.
Requirements
— For external adjustment: The "EXT.CAL." configuration value is set in the
"SETUP " / "BALANCE" / "CAL.ADJ." menu.
— For the adjustment function based on calibration and automatic
adjustment: The "CAL/ADJ." configuration value is set in the
"SETUP " / "BALANCE" / "CAL.SEQ." menu.
— The weighing pan is unloaded.
— The displayed weight value remains unchanged.
— The device is not locked against external calibration. Devices with special
country approvals may be locked against external calibration.
— The adjustment weight is already there.
Procedure
Press the [Zero] key.
t
If the device cannot be zeroed: Remove the sample to be weighed and
t
re-zero the device.
Press the [Adjust] button.
t
If automatic leveling is configured: The device levels itself automatically.
y
Place the indicated calibration weight on the balance.
t
The calibration and adjustment function is prepared:
y
— If the adjustment weight placed on the device is within the specified
limits: The adjustment process can be started.
— If the weight placed on the device is too high: The "+" symbol is
displayed.
— If the weight placed on the device is too low: The "-" symbol is
displayed.
Press the [Confirm] button. This starts the calibration.
t
The [CAL.END.] message indicates the end of calibration.
y
Remove the calibration weight after calibration is complete.
t