System Settings
52
Operating Instructions Cubis
7.3
Entering the License Key
For the licensing of device functions, the corresponding application must be
stored on the device. The application may be available upon delivery of the
device, or it may saved to the device at a later stage by Sartorius Service.
Requirements
The application is stored on the device and the license key is known.
Procedure
Open the INPUT" / „QAPP.LIC. / QAPP" menu.
t
Enter the designation for the licensed device functions, e.g. "QP10"
t
The "LICENSE" menu is opened.
y
Enter the 8-digit license key.
t
7.4
Switching Off the isoCAL Function
If the isoCAL function is switched off for a conformity-assessed device: The
device can only be used in restricted temperature ranges for legal-for-trade
applications (see Chapter "15.3 Ambient Conditions", page 78). The
isoCAL function cannot be switched off for all model versions.
Procedure
In the "SETUP" / "BALANCE" menu, for the "ISOCAL" parameter, select
t
the "OFF" configuration value.
7.5
Activating, Deactivating or Configuring the
Ionizer
Requirements
A license key has been entered to use the ionizer.
Procedure
NOTICE Risk of damage to the power supply unit from argon! Observe
t
the instructions for using argon (see Chapter "15.7 Ionizer when Using
Device in an Isolator with Protective Argon Atmosphere", page 80).
To deactivate the ionizer: In the "DEVICE" / "IONIZER" / "POWER" menu,
t
select the "OFF" setting.
To activate the ionizer: In the "DEVICE" / "IONIZER" menu, set the
t
desired intensity and duty cycle of the ionization process.
Exit the settings menu.
t
The [Ionizer] button appears in the operating display.