Troubleshooting Checklist
Table 7-1. Troubleshooting Checklist
Symptom
Patient
■
uncomfortable,
feels hot
sensation
English 8930 System User Guide 2010-12
MA07445A008
Rev X
Probable Cause
Full bladder
Empty bladder.
■
■
High urethral
Turn on irrigation
■
■
temperature
for a few seconds.
Medtronic Confidential
Actions
Prevention
Be sure bladder is
■
appropriately
drained.
Maintain adequate
■
shielding. Maintain
adequate
analgesic
/anesthetic control.