Shutting Down System, Disconnecting Cables And Tubing System - Medtronic Prostiva 8930 Guía Para El Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 383
8930EN_4.fm 12/17/10 2:44 pm
UCxxxxxx EN
7 x 9 inches (178 mm x 229 mm)

Shutting Down System, Disconnecting Cables and Tubing System

MA07445A008
Rev X
When the patient's treatment is completed, or there is a safety
shutdown, and no further use of the system is required, complete
the following steps to disconnect the cables and tubing system:
1. If not previously done, carefully remove the Model 8929
Hand Piece from the patient.
Warning: Do not continue, reposition, or remove the
hand piece until you have confirmed that the trigger
mechanism is fully forward. Failure to place the trigger
mechanism into the needles-fully-retracted position can leave
the needles and shields partially deployed. This can result in
possible patient injury if the hand piece is repositioned or
removed from the urethra.
2. Turn off the Model 8930 Radio Frequency Generator system
power.
3. Remove the Model 8943 Return Electrode Cable from the
radio frequency generator by pulling straight out on its
connector plug. Carefully remove the return electrode from
the patient by gently peeling it back from the patient.
4. Remove the Model 8929 Hand Piece radio frequency cable
from the generator by pulling straight out on its connector
plug.
5. If installed and no additional treatments are scheduled for this
use, remove the Model 60883 Remote Foot Switch cable by
grasping release mechanism on the connector and pulling
straight outward to disengage and remove.
6. If no additional treatments are scheduled for this use, remove
the power cord from the wall socket outlet; then remove the
power cable from the back of the generator and store it with
the generator.
7. Disconnect the user-supplied light cable from the telescope,
and remove the telescope from the Model 8929 Hand Piece.
Note: The telescope can be cleaned and resterilized in
accordance with the instructions in the telescope instructions
for use. The Model 8929 is a single-use only device and it
cannot be reuse.
8. Close the Model 6101 Tubing System thumb-roller clamp and
disconnect the tubing system from the hand piece.
9. Discard the Model 8929 Hand Piece, Model 8934 Return
Electrode, and Model 6101 Tubing System in accordance
with local environmental regulations.
Prostiva RF Therapy Procedure
Patient Treatment
2010-12 8930 System User Guide English
Medtronic Confidential
CRMRef_R06
81

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Prostiva 8929Prostiva 8099Prostiva 8099tu15

Tabla de contenido