Patient and Hand Piece Preparation

Setting Up Hand Piece

English 8930 System User Guide 2010-12
MA07445A008
Rev X
Complete the instructions outlined below to configure the Model
8929 Hand Piece as shown in the (Figure 4-8) illustration.
Spring-loaded
Syringe
Two-way
Stopcock
Light Source
Cable
Connector
Telescope Lock
Telescope
Figure 4-8. Set Up Hand Piece and Attached Accessories.
1. Use sterile technique to open the Model 8929 Hand Piece
sterile tray and place the hand piece into the sterile field.
Caution: The hand piece useful life and sterility may be
compromised if the device is removed from the sterile
packaging or connected to the RF generator more than
30 minutes prior to starting the procedure.
2. Open and connect the sterile Model 6101 Tubing System.
a. Check that the thumb-roller clamp is closed.
b. Attach the irrigation source tubing to the tubing system at
the Luer-lock fitting that is connected to the thumb-roller
clamp.
c. Attach the tubing from the Y-connection to the hand piece
two-way stopcock.
Medtronic Confidential
Irrigation Source
Luer-lock Fitting
Thumb-roller Clamp
Y-Connection
Model 6101Tubing System
RF On/Off Switch
Model 8929 Hand Piece