Medtronic Prostiva 8930 Guía Para El Usuario página 44

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 383
8930EN_2.fm 12/17/10 2:43 pm
UCxxxxxx EN
7 x 9 inches (178 mm x 229 mm)
Chapter 2
44
User Interface
English 8930 System User Guide 2010-12
MA07445A008
Rev X
Table 2-7. Alert Messages (continued)
No.
Message
17
Return electrode not
connected
Connect return electrode
to patient and generator
18
Return electrode problem
Replace with split return
electrode supplied by
Medtronic
19
Impedance problem
Check hand piece and
return electrode
Medtronic Confidential
CRMRef_R06
Explanation
Return electrode is not
properly applied to the
patient, or the cable is not
connected to the radio
frequency generator
Return electrode must be
connected in order to
proceed
Message stays on until
user properly applies
return electrode or
connects cable
Alarm is 2 short beeps
+ 1 long beep if radio
frequency on; if off,
2 short beeps
Return electrode being
used is not compatible
Replace return electrode
with Medtronic split-pad
Message stays on until
user replaces return
electrode
Alarm is 2 short beeps
Radio frequency circuit is
not complete, or has a
problem (open > 999
ohms; short <35 ohms)
Check cable connections
Message stays until user
corrects and impedance is
no longer out of range;
displays for a minimum of
ten seconds
Alarm is 2 short beeps
+ 1 long beep if radio
frequency on; if off,
2 short beeps