Medtronic Prostiva 8930 Guía Para El Usuario página 43

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 383
8930EN_2.fm 12/17/10 2:43 pm
UCxxxxxx EN
7 x 9 inches (178 mm x 229 mm)
MA07445A008
Rev X
Radio Frequency Generator Description
Table 2-7. Alert Messages (continued)
No.
Message
14
Hand piece problem
Replace hand piece
15
Usage limit reached
Replace hand piece
16
Hand piece switch
problem
Check hand piece switch
or use foot switch
User Interface
Explanation
Hand piece is not properly
functioning
Hand piece must be
replaced with new unit in
order to proceed
Message stays on until
user replaces hand piece
Alarm is 2 short beeps
+ 1 long beep if radio
frequency on; if off,
2 short beeps
Hand piece is at its
maximum-use limit; it can
no longer be used
Hand piece must be
replaced with new unit in
order to proceed
Message stays on until
user replaces hand piece
Alarm is 2 short beeps
Hand piece switch is stuck,
or kept pressed for more
than five seconds
Hand piece switch should
be checked
Message stays on until
user takes corrective
action or disconnects the
hand piece
Alarm is 2 short beeps
+ 1 long beep if radio
frequency on; if off,
2 short beeps
2010-12 8930 System User Guide English
Medtronic Confidential
CRMRef_R06
43

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Prostiva 8929Prostiva 8099Prostiva 8099tu15

Tabla de contenido