Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 383
8930EN_1.fm 12/17/10 2:43 pm
UCxxxxxx EN
7 x 9 inches (178 mm x 229 mm)

Tubing System

MA07445A008
Rev X
Overview of the Prostiva Therapy System
On the hand piece, there is a telescope slot assembly that
secures the telescope in position. This assembly also moves 1-2
cm to provide repositioning of the telescope to adjust the view.
Figure 1-3. Telescope
The telescope is supplied as a non-sterile medical device. It is
reusable and must be cleaned, sterilized, and used in accordance
with the applicable instructions for use.
The Model 6101 Tubing System connects to the Model 8929
Hand Piece to supply cooling fluid during the Prostiva RF Therapy
procedure (Figure 1-4). This system connects to an irrigation
source. The system helps provide clear visualization and any
required cooling of the urethra. If the urethra temperature reaches
°
43
C, a caution message to start irrigation appears on the
screen. If the urethra temperature reaches 47
automatically stops, and an alert message appears on the screen.
Accessories
°
C, the lesion
2010-12 8930 System User Guide English
Medtronic Confidential
CRMRef_R06
17