Radio Frequency Generator Description
Table 2-6. Status Messages
No.
Message
1
System test in progress...
2
System ready
Start new lesion
3
Lesion in progress....
User Interface
Explanation
Radio frequency generator
system test of internal
electronics and attached
accessories (ie, hand
piece, return electrode) is
running
Message stays on until
system test is completed
(approximately five
seconds)
No alarm occurs during
system test
Radio frequency generator
and attached accessories
(eg, hand piece, return
electrode) are ready for
use
Message stays on until
clinician starts radio
frequency, or an alert
occurs
No alarm occurs during
system ready
Both blue LEDs flash
continuously at a regular
rate
Lesion treatment with both
needles has started and is
continuing
Message stays on during
lesion unless caution
message displays
"Lesion Start" alarm is
1 medium beep
"Radio Frequency On"
alarm is 1 short beep
every 4 seconds
Both blue LEDs are on
solid