Chapter 2
42
User Interface
English 8930 System User Guide 2010-12
MA07445A008
Rev X
Table 2-7. Alert Messages
No.
Message
11
Generator problem
Restart generator
Contact Medtronic if
problem persists
12
Generator overheated
Check air vents and wait 5
minutes before
continuing
Contact Medtronic if
problem persists
13
Hand piece not connected
Connect hand piece to
generator
Medtronic Confidential
CRMRef_R06
Explanation
Radio frequency generator
■
has a hardware or software
problem that may be
recoverable
Problem may be
■
correctable if user restarts
system
Message shows briefly
■
until system displays a
solid blue screen; this may
clear when system is
restarted
Alarm is 2 short beeps
■
+ 1 long beep if radio
frequency on; if off,
2 short beeps
Radio frequency generator
■
is overheated
Check bottom vents and
■
fan clearances; allow
generator to cool for at
least five minutes
Message stays on for five
■
minutes until generator is
cool enough to proceed
Alarm is 2 short beeps
■
+ 1 long beep if radio
frequency on; if off,
2 short beeps
Hand piece cable is not
■
properly connected to the
radio frequency generator
Hand piece must be
■
connected in order to
proceed
Message stays on until
■
user connects hand piece
cable
Alarm is 2 short beeps
■
+ 1 long beep if radio
frequency on; if off,
2 short beeps