While calibrating pipettes use pipettes database providing you with pipette related data,
parameters, tested volume, and error values determined for specific volumes.
Prior pipette calibration process, enter pipettes database and provide pipette-related data and
calibration criteria. In order to add pipette it is necessary to enter the database. For information on
how to add pipette to pipettes database read down this user manual.
Pipettes Calibration mode screen

24.1. Mode Related Settings

The supplementary settings enable you to adjust the working mode to your needs and requirements.
To access the settings follow the below procedure.
Procedure:
1. Press grey workspace area.
2. The following submenus are displayed: Settings, Buttons, Information, Printouts, Profile.
3. Select <Settings> menu.
4. The display indicates functions related to pipette calibration process.
Pipettes Calibration mode settings:
Measurements quantity: designed to enable determining measurements quantity for tested
pipette volume (valid for each tested volume in case of a pipette with adjustable volume),
Request serial number: YES/NO. YES – a window with an on-screen keyboard is displayed,
use it to enter serial number of a pipette.
Operation with adherence to ISO 8655: YES/NO. YES – the software automatically takes
error values accordant with recommendations of ISO 8655 standard (if other error values
have been defined for a particular pipette, then on selecting <Operation with adherence to
ISO 8655> parameter the specified error values are neglected and replaced by those from
ISO standard).
Upload ambient conditions parameters from THB module: YES/NO. YES – the software
automatically acquires and saves ambient conditions' parameters from a connected THB
sensor at a right time. NO – you have to read ambient conditions (i.e. temperature, humidity
and pressure) from external sensors and enter them manually both before and after
calibration.
Result control, Tare mode, Automatic footer printout, Printout mode, Printout: for
information on how to operate these functions read section 15.7 .
24.2. Pipettes Calibration – Quick Access Buttons
Each working mode features automatically displayed set of default on-screen quick access buttons.
The set can be modified, i.e. you can add out-of-list quick access buttons to it. Such operation
requires particular permissions level.
While using automatic chamber for pipettes calibration you can control the opening for liquid dosing,
this is done either via quick access buttons or proximity sensors:
- 103 -
1. Setup – press to access
balance menu.
2. Print header.
3. Print footer.
4. Select pipette.