Pipettes calibration procedure:
1.
Press <
Select pipette> button.
2.
List of pipettes is displayed, select pipette to be calibrated.
3.
Pipettes calibration home screen is displayed, selected pipette name is visible in Info field.
4.
Press <
START> button.
5.
With <YES> value set for <Request serial number> parameter, a window for entering the
number opens. Enter the value and confirm.
6.
Window for entering ambient conditions values is displayed, read the temperature, humidity and
pressure from external sensors and enter them, next press <
With <YES> value set for <Upload ambient conditions parameters from THB module>
parameter, the temperature, humidity and pressure values are acquired automatically from the
THB module, water temperature must be entered manually, enter it and press <
confirm.
7.
Pipettes calibration home screen is displayed, tested volume and process status <In progress>
are displayed in Info field. The task bar displays prompts guiding you through the process -
<C1/V1/N1 Sample weighing>.
Where: C1 – channel number; V1 – volume number for the channel; N1 – measurement number
for tested volume.
8.
Keep
following
Info field provides you with information on the on-going process (average volume, error values).
9.
Upon confirmation of the last measurement, a window with ambient conditions parameters is
displayed again. Enter the parameters and press <
10. A report is generated automatically, the report is printed and recorded in pipettes calibration
reports database (ambient conditions and water temperature values given on a report are
average values from the process);
11. Press <
> button located in the bottom bar. For multichannel pipette you are asked whether
the process shall be continued for remaining channels. Upon confirmation the balance carries
out calibration process for the following channels. The settings remain unchanged.
12. You can start the following procedure for the same pipette or select another pipette form pipettes
database.
24.6. Performed Pipettes Calibrations Report
Upon completed calibration process a report is generated. The report is saved to the <
calibration reports> database. The files are named by date and time (hour) of calibration process
performance.