Press to print the graph using PCL printer.
Press to save the graph as *.bmp file to USB flash drive connected to USB port.
24.

PIPETTES CALIBRATION

Balance enables pipettes calibration using dedicated program function or PIPETTES PC software
(workstation for pipettes calibration). Prior to calibration start it is necessary to install pipettes
calibration set inside the anti-draft chamber. The set does not come standard with the balance.
MYA 4Y balance with a set for pipettes calibration.
Installation procedure:
1. disassemble the standard weighing pan and the anti-draft shields,
2. inside the chamber install:
draft shield (2),
weighing pan (3),
glass vessel (5),
evaporation ring (6),
glass lid (7),
supplementary lid (8), or
glass lid (4).
- 98 -