Remember!
Total weight value of all parts loaded onto the weighing pan cannot be greater than the max capacity
value.
Total weight value of all parts loaded onto the weighing pan cannot be lower than the value determined
by <Minimum reference sample mass> parameter. Unless this condition is met, the weighing
instrument displays a message: <Sample mass too low>.
Single part mass value must be equal to or greater than 0.1 of the reading unit. Unless this condition is
met, the weighing instrument displays a message: <Single part mass too low>.

16.5. Setting Reference Sample Mass by Acquiring Mass of a Single Part from Database

Each product recorded in a database of products features set of information which allows to identify
it. Among the information there is weight value used during part counting process.
Procedure:
Enter <
Parts counting> mode and press <
product from the list.
Entering reference sample mass to weighing device memory
Procedure for entering single part mass to products database:
1. Press <Setup> button, next press <Databases> entry and select Products database.
2. Press <Products> button.
3. Press respective product name and modify <Mass> field, [5].
4. Go back to <Parts counting> mode.
Procedure for entering single part mass to products database when there is no data entered yet:
a) Press <Setup> button, next press <Databases> entry and select Products database.
b) Press <
c) Fill in or product-related fields and modify <Mass> field, [5].
d) Go back to <Parts counting> mode.

16.6. Single Part Mass Update in a Database

Determined single part mass can be assigned to a product in the database of products. This option is
applied while using ACAI (Automatic Accuracy Correction) function in order to determine part mass
with high accuracy.
Procedure:
1. Determine single part mass.
2. Press <
Products> database.
3. Press and hold name of a product for which mass value is to be updated.
4. Context menu openes.
5. Press <Assign reference sample mass> button, reference sample mass for the product is
recorded as <Mass>.

16.7. Parts Counting Procedure

First, it is necessary to obtain weight value of a single piece. Select one of the available options:
Add> button.
Products> database pictogram. Select respective
- 75 -