Example of settings modification:
For better understanding, analyse the following example (the values used below are exemplary ones,
they were not taken form the balance).
1. Assumed vibrations threshold set by default on 4Y series balance: 100mm/s
2. Max vibrations value: 100% (threshold for alert display – pictogram colour change, from green to
red). Pictogram colour is green for vibrations within 0 - 100mm/s
2
101mm/s
, the pictogram changes to red.
3. Max vibrations value: 150%. Pictogram colour is green for vibrations within 0 - 150mm/s
for vibrations value of 151mm/s
Vibrations are signalled by a respective pictogram displayed in the weighing result window.
Upon pressing vibrations pictogram, window with current values measured by sensors is displayed.
Displayed vibrations value (IS V) is the effective value registered within the most recent 2 seconds.
2
, the pictogram changes to red.