34.9. Level Control
The balance is equipped with Automatic Level Control mechanism. You can define its operation for
non-verified devices.
In case of verified instruments, the settings are hidden and the balance operates in accordance with
factory settings, i.e. <Active with lock>. Weighing process is possible only if balance is properly
levelled.
Procedure:
1. Enter <
Misc.> submenu.
2. Select <
Level control> parameter, an edit box is displayed.
3. Select one of available options:
•
None – level indicator not displayed, the balance does not control level status,
•
Active – level indicator displayed, level status is signalled by colour (green→ OK, red→
levelling required),
•
Active with lock – level indicator displayed, level status is signalled by colour (green→ OK,
red→ levelling required). For red level indicator 'no Level' alert is displayed - weighing
function is blocked.
Caution! For detailed information on levelling read section 10.3 of this user manual.
34.10.
Decimal Point
Parameter allowing to set punctuation mark for a printout.
Procedure:
1. Enter <
Misc.> submenu.
2. Select <Decimal mark> parameter, an edit box is displayed.
3. Select one of available options:
a. Dot
b. Coma
Submenu screen is displayed.
34.11.
Proximity Sensors Sensitivity
Parameter allowing you to change sensitivity of the proximity sensors, the values range
between 0 - 9.
By default it is 5 - 7.
Procedure:
1. Enter <
Misc.> submenu.
2. Select <Proximity sensors sensitivity> parameter, an edit box is displayed.
3. Select one of available options.
34.12.
Proximity Sensors Timeout
Parameter for setting timeout of proximity sensors expressed in [ms].
Default value: <0>.
When you use the option of motion recognition, set <Proximity sensors timeout> parameter to
<500ms> value.
Procedure:
1. Enter <
Misc.> submenu.
2. Select <Proximity sensors timeout> parameter, an on-screen keyboard is displayed.
- 171 -