Copy Tare - RADWAG 4Y Serie Manual Del Usuario

Ultaramicrobalanza. microbalanza. microbalanzas para pipetas. microbalanzas para los filtros. balanzas analíticas. balanzas analíticas para los filtros
Ocultar thumbs Ver también para 4Y Serie:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Weighing B
Upon pressing Weighing B button, the balance searches samples list in order to find the first
sample for which weighing <B> is still to be carried out. If there are no such samples the
balance displays a message informing you that the operation is disabled. If the operation is
enabled then new data on initiated process is displayed in the grey workspace.
Caution! Next to <Weighing B> caption you can see digits 1/3 where <1> means the first
measuring cycle of type B, and <3> means that <Repetitions Quantity> parameter is set to <3>
value (3 cycles).
The software prompts you to carry out the <First> cycle, out of <3> cycles, for all samples in the
series for which the measurement is enabled (it is required that weighing A is already completed
for the sample).
You are guided through the process by means of prompts visible either in the workspace or
bottom bar.
If tare is assigned to a sample, then its value is displayed with '-' sign. Load the weighing
pan with a sample (if tare is assigned load the sample in a container) and press <
For <Ambient conditions> parameter set to <Value> option, a window for entering
temperature and humidity values is displayed. For <Ambient condition> parameter set to
<Online> option, a window for confirmation of values taken from ambient conditions module
is displayed.
Caution! If <Ambient conditions> parameter for the differential weighing mode is set to <No>
value the software skips this part of the process.
Press <
> button to confirm ambient conditions values, differential weighing home
screen is displayed, you are prompted to unload the weighing pan.
Unload the weighing pan and press <
again in order to find the first sample for which weighing <B> is still to be carried out. Follow
the above procedure and carry out weighing of the remaining samples.
In order to abort the process press <
Upon completing measurements of final mass for the samples in a series, you can check the results
in the series database. To check the results, enter the database, select a series, then select samples
and a specific sample for which <Weighing B> has been carried out.
Particular weighing's preview mode provides you with values. Numbers referring to weighing <B>
inform on the measuring cycle.
Data on a specific series can be printed or exported to a file.

25.5. Copy Tare

Copy Tare option enables copying once selected tare value (in a series) to all samples which do not
have tare value assigned, and for which differential weighing process (status other than Weighing B)
has not been completed.
> button. The balance searches samples list
> button.