10.2. External Adjustment

External adjustment is carried out using external mass standards of the right accuracy and weight
value, which value depends on balance model and capacity. The external adjustment procedure is a
semi-automatic process, particular stages of which are signalled with on-screen prompts.
Caution! External adjustment is possible for balances that are not a subject to the conformity
assessment (verification).
Procedure:
1. Enter <Adjustment> submenu, next select <
2. Message ordering you to remove weight is displayed. Unload the weighing pan and press
button. The balance determines start mass, message: Start mass determination is displayed.
3. Upon completed start mass determination, message ordering you to load the weighing pan is
displayed. Load the pan with weight of respective mass value, and press
4. In case of some balances the next message is displayed, following it load the weighing pan with
the new mass and press
5. Upon completed procedure unload the weighing pan, press
returns to the weighing mode.

10.3. User Adjustment

User adjustment can be carried out using any adjustment weight, mass of which ranges between
0.15 Max and Max. User adjustment and external adjustment procedures are likewise with one
exception, before user adjustment start, an edit box for entering mass of the used adjustment weight
is opened.
Caution! User adjustment is available for balances that are not a subject to the conformity assessment
(verification).
In order to start user adjustment, enter <Adjustment> submenu and select <
option. Then follow commands displayed on the screen.

10.4. Adjustment Test

<Adjustment test> function enables comparing the result of an internal automatic adjustment with the
value recorded in balance factory parameters. The comparison is used for determining drift of
balance sensitivity over time.

10.5. Automatic Adjustment

Enter this menu to specify conditions initiating an automatic adjustment. Options:
None – automatic adjustment inactive.
Time – adjustment takes place in time intervals declared in <Automatic adjustment time>
menu (10.6).
Temperature – adjustment is triggered by temperature change.
Both – adjustment is triggered by both, temperature changes and time.
Caution! This parameter settings can be modified only for balances that are not a subject to the
conformity assessment (verification).

10.6. Automatic Adjustment Time

<
Automatic adjustment time> parameter determines time interval between successive automatic
internal adjustments. The time interval is declared in hours and ranges between 1 and 12 hours.
To set time interval for automatic adjustment:
enter <Automatic adjustment time> parameter,
External adjustment> parameter.
button.
- 41 -
button.
button to confirm, the balance
User adjustment>