using displayed menu, select appropriate time interval (given in hours) which is to elapse
between the most recent adjustment and activation of the following automatic internal
adjustment.
Caution! This parameter settings can be modified only for balances that are not a subject to the
conformity assessment (verification).
10.7. Scheduled Adjustments
<
Scheduled adjustments> parameter allows you to precisely specify time for adjustment
performance and interval between successive adjustments. The parameter is not connected with
automatic adjustment and its criteria (time, temperature). You can schedule when internal and
external adjustment are to be performed. In order to design schedule for external adjustment, it is
necessary to record mass standards for adjustment performance into balance memory.
Procedure:
1. Enter data of mass standards intended for external adjustment performance; to do it:
2. Enter operator menu, select <Adjustment> submenu and find <
parameter.
Enter operator menu, select <Adjustment> submenu and
find <Scheduled adjustments> parameter.
Enter <Scheduled adjustments> parameter, wait for
'Scheduled adjustments' window to open, add adjustment
procedures. Remember, only Administrator can add new
entries.
- 42 -
Adjustment weights>