module to balance ports COM 1 or COM 2 and set appropriate communication parameters, for the
parameters refer to a data plate.
15.7. Additional Parameters for a Weighing Process
Modification of weighing functions enables programming balance operation.
Procedure:
1. Press grey workspace area.
2. The following submenus are displayed: Settings, Buttons, Information, Printouts, Profile.
3. Enter particular submenu and select the component that is to be modified.
<Settings> - additional options related to weighing mode,
<Keys> - defining quick access buttons,
<Information> - supplementary data on weighing process displayed in an Info field,
<Printouts> - selecting printout type,
<Profile> - selecting profile that is to be active during balance operation,
<SETTINGS> - weighing process related data such as:
Tare mode
•
SINGLE
Mass value saved to balance memory on single pressing of TARE key. Subsequent pressing
of TARE key determines new tare value. Selecting product or packaging to which tare value
is assigned causes automatic deleting of previously assigned tare value.
•
CURRENT SUM
Total value of currently entered product and packaging tare values (result of selection of
product and packaging from the database) with possibility to add tare value entered
manually. When new tare value for a product or packaging is set again then tare value
entered manually gets deactivated.
•
TOTAL SUM
Total value of all entered tare values.
•
AUTOTARE
Means of operation:
Each first stable measurement result is tared. NET sign is displayed. Now you can
determine net mass of weighed load. Upon unloading the weighing pan and autozeroing,
tare value is automatically deleted.
Automatic footer printout
•
MODE
None – manual footer printout,
Sum of measurements – footer is printed when mass value entered in <Threshold>
parameter gets exceeded.
Measurements quantity – footer is printed when a pre-defined measurements quantity (a
batch) set in parameter <Threshold> is carried out.
•
THRESHOLD
Setting the threshold value determining footer printout.
For <Sum of measurements> parameter, threshold value is determined in measuring unit
[g].
For <Measurements quantity> parameter, the unit is not determined, the provided value
specifies carried out measurements quantity.
Printout/Enter mode
•
PRINT/RELEASE button (manual operation)
Never
First stable
Each stable
Each
•
AUTOMATIC MODE
Never
– printout disabled.
– first stable measurement is recorded.
– all stable measurements are registered.
– printout of all measurements (stable and unstable), for verified
balance only stable measurements are printed out (as in <Each
stable> setting).
– printout disabled.