24.3. Adding a Pipette to Pipettes Database

Database of pipettes contains list of pipettes' names and other data, including tested volumes and
error values for a specific volume. When adding a pipette to the database, first specify its name, and
then add remaining data.
The software is intuitive and you are guided by means of respective messages. Pipettes database
menu has been designed to enable adding formulations.
Procedure:
1. Enter <
Databases> submenu and press <
2. Press <
Add> button to add a new pipette.
A new record is automatically added to the database, it is opened in an editing mode. Enter
respective data.
List of parameters defined for a pipette:
Name: press to enter pipette name.
Code: press to enter pipette code.
Model: press to enter model name.
Tip: press to enter tip name
Volume type: ADJUSTABLE/FIXED.
Nominal volume: pipette's nominal volume value.
Minimum volume: pipette's minimum volume value, for fixed volume pipette enter <0>;
Channels quantity: specify number of pipette's channels, for a single channel pipette enter
<1>;
Type: NONE/A/D1/D2. Pipette type adheres to standard guidelines. In case of calibration
carried out with reference to ISO 8655 standard the errors are automatically selected for
particular volume. Determining pipette type is crucial.
Tested volumes: press "Tested volumes" field, a window with a list of calibrated volumes
opens (in case of a new pipette the list is empty), add volume to be tested, and define error
value for each tested volume:
Procedure:
1. Press <
2. An on-screen keyboard is displayed.
3. Enter tested volume value in [µl] and press <
4. A new record is automatically added, the record features recommended error values.
In order to change error values press a field with pipette volume (each field can be edited,
you can enter error values of your choice).
Caution! During calibration, the order of tested volumes corresponds to the order in which they
are entered.
5. Upon setting correct values return to the home screen.
Add> button.
Pipettes> button.
> button to confirm,