Prior to each use, the NVM5
Control Unit Inspection
Inspect the touch screen for damage and confirm that all labels are legible and in good
condition.
Confirm that the power cord insulation is in good condition and that there are no exposed
wires.
If any damage is observed, the device must be taken out of service and returned to
NuVasive
.
®
Patient Module Inspection
Insert the Patient Module cable connector into the corresponding Control Unit receptacle
and confirm that it seats fully.
Confirm that the cable insulation and Patient Module housing are in good condition and that
there are no exposed wires.
If any damage is observed, the device must be taken out of service and returned to
NuVasive.
Camera Inspection
If the Camera has suffered damage or an impact, which has affected its calibration, the
amber colored LED will appear on the Camera face, and / or alert message will appear on
the Bendini
software.
®
The software message will state "Camera communication error. Requires calibration."
If this error is observed, please contact NuVasive Customer Service to coordinate
equipment servicing.
Electrical Safety Test
Upon initial receipt, and every 12 months thereafter, the NVM5 System should be inspected
and an Electrical Safety Test should be performed, including leakage current testing, to assure
that the device continues to comply with applicable electrical safety standards for hospital grade
medical devices.
Preventive Maintenance
It is strongly recommended that the NVM5 System undergo factory inspections and
maintenance procedures every 12 months. Contact NuVasive Customer Service at 1(800) 475-
9131 to arrange preventive maintenance.
Service and Repair
The NVM5 System has no user serviceable parts inside. Servicing and repair is to be
performed only by NuVasive or its authorized agent.
Disposal of Equipment and Accessories
The disposable leads and accessories used with the NVM5 System are only intended for
single-use, and must be discarded in accordance with standard hospital practices for controlling
medical waste.
MAINTENANCE
System should be inspected and tested as follows:
®
12