TECHNICAL SPECIFICATIONS cont.
Guidance and manufacturer's declaration—electromagnetic emissions
The NVM5
System is intended for use in the electromagnetic environment specified below. The customer or the
®
user of the NVM5 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration—electromagnetic immunity
The NVM5 is intended for use in the electromagnetic environment specified below. The customer or the user of the
NVM5 should assure that it is used in such an environment.
Immunity test
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±1 kV for power
Transient/burst
supply lines
IEC 61000-4-4
±0.5kV for
input/output lines
Surge
±1 kV differential mode
IEC 1000-4-5
±2 kV common mode
Voltage dips, short
<5 % U
Interruptions and
(>95% dip in U
voltage variations
for 0.5 cycles
on power supply
Input lines
<40 % U
(60% dip in U
for 5 cycles
IEC 61000-4-11
<70 % U
(30% dip in U
for 25 cycles
<5 % U
(>95% dip in U
for 5 s
Power frequency
30 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE - U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
The NVM5 System uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
Group 1
not likely to cause any interference in nearby electronic
equipment.
The NVM5 System is suitable for use in all establishments,
Class A
including domestic establishments and those directly
connected to the public low-voltage power supply network that
Class A
supplies buildings used for domestic purposes.
Complies
IEC 06901
Compliance level
test level
±6 kV contact
±8 kV air
±1 kV for power
supply lines
±0.5kV for
input/output lines
±1 kV differential mode
±2 kV common mode
<5 % U
T
)
(>95% dip in U
T
for 0.5 cycles
<40 % U
T
)
(60% dip in U
T
for 5 cycles
<70 % U
T
)
(30% dip in U
T
for 25 cycles
<5 % U
T
)
(>95% dip in U
T
for 5 s
30 A/m
30
Electromagnetic environment--guidance
Electromagnetic environment—
Floors should be wood, concrete,
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
T
)
of a typical commercial or hospital
T
environment. If the user of the
NVM5 requires continued operation
during power mains interruptions,
T
)
it is recommended that the NVM5
T
be powered from an uninterruptible
power supply or a battery.
T
)
T
T
)
T
If image distortion occurs, it may
be necessary to position the NVM5
further from sources of power
frequency magnetic fields or to
install magnetic shielding. The
power frequency magnetic field
should be measured in the intended
installation location to assure that it
is sufficiently low.
guidance