Mount Camera onto stand or IV pole using
attached locking clamp. Ensure Camera is
mounted at the height of the Array during
digitization.
Camera must be at a minimum of 22 inches
(1.83 feet), and no more than 52 inches (4.33
feet) away from the patient. The Camera should
be orthogonal to the long axis of the patient's
spine.
Plug Control Unit into wall outlet; ensure power
is clean and grounded. Press Power Button to
turn System on.
1. Connect Camera Cable to connection on back of Camera. Line up the red dots on the
cable and the Camera to ensure a proper connection.
2. Connect USB Cable to any USB Port Connection on Control Unit.
3. Connect Power Supply Cable to a grounded outlet.
Camera Cable
EQUIPMENT SETUP cont.
®
2. USB Cable
3. Power Supply
Cable
3
Camera
Control Unit USB
1. Camera Cable