TECHNICAL SPECIFICATIONS cont.
SSEP Stimulation Output
Waveform:
Polarity:
Output Regulation:
Pulse Width:
Current Pulse Amplitude:
Stim Rate:
Maximum Voltage:
SSEP Monitor
Number of Channels:
Frequency Response:
A/D Converter:
Full Scale Span:
A/D Sampling Rate:
Common Mode Rejection Ratio:
Bendini
Power Requirements
®
Voltage:
Frequency:
Current:
Classifications
•
The NVM5
System is a Class l, Type BF Applied Part device with respect to protection
®
against electric shock.
•
The NVM5 System is considered Ordinary Equipment – IPX0 – in that it is not
protected against water ingress.
•
The NVM5 System is not suitable for use in the presence of flammable anesthetic
mixtures.
•
The NVM5 System is classified suitable for Continuous Operation.
•
Only equipment and accessories complying with EN 60601-1 and approved for use
with the NVM5 System may be connected.
•
The Bendini System is a Class l Device with respect to protection against electric
shock.
•
The Bendini System is considered Ordinary Equipment – IPX0 – in that it is not
protected against water ingress.
•
The Bendini System is not suitable for use in the presence of flammable anesthetic
mixtures.
•
The Bendini System is classified suitable for Continuous Operation.
Rectangular, Monophasic Pulse
Polarity Selectable
Constant Current
50 to 300 microseconds
1 to 100mA (dependent on electrode impedance)
1.7 Hz to 5.7 Hz
300V
7
30Hz to 300Hz
16 bit resolution
± 2,048 mV
9.6 kHz
Greater than 74dB
100-240 VAC
50-60Hz
1.2-0.63 A
28