Monitor Case Status
Total
Last Injection
Volume
Volume
Selected
Threshold
Note: graphical range indicator is not representative of a recommended
contrast dosage
The Pressure Module connects wirelessly to the Monitor.
It includes a module with a pause button and a pressure
transducer.
The Smart Syringe and Pressure Module have two LEDs
to provide user feedback.
Solid green (active for 10 seconds) indicates
injections will be counted towards cumulative volume
Flashing Yellow indicates system is paused and
injections will not be counted towards cumulative volume
Assemble and Prime
Note: It is recommended to leave all battery pull tabs in the disposable devices
until all priming is complete and the case is ready to begin. This will
ensure maximum battery life and performance
throughout the case.
1) Introduce disposables into sterile field per
normal procedure.
2) Attach contrast source line to the female luer located in the
pressure transducer of the Pressure Module.
3) Attach the male luer of the pressure
transducer to the Manifold, and the Manifold
to the Smart Syringe.
4) Prime manifold and all lines per normal procedure.
Wirelessly Connect Disposables to Monitor
Note: Assemble and prime system prior to wirelessly connecting disposables
1) Remove battery pull tab battery pull tab from the Smart Syringe by pulling
the tab straight out from the syringe plunger. Remove battery pull tab from
the Pressure Module by pulling the tab straight out from the Module (i.e.
perpendicular from the back of the Module). Do NOT hold the Pressure
Module down against the table such that the pull tab is against the table
when removing the pull tab.
2) Select New Case. Wirelessly connect Smart Syringe & Pressure Module to
the Monitor by selecting the device to connect on the display.
Contrast
% Mix or % Saved
Volume Remaining
Until Selected
Threshold Reached
Or
Pause
Button
Pressure
Transducer
Female Luer
(Contrast
Source)
Male Luer
Male Luer
(Manifold)
(Manifold)
3) Confirm wireless connection of the Display and disposables. Following
scanning, the LED's on the Smart Syringe & Pressure Module will flash in
the same patter as shown on the Display for proper identification.
a) Select Yes on the Display or move the plunger of the Smart Syringe >
2 mL to confirm Syringe flashing.
b) Select Yes on the Display or press Pause Button on Pressure Module to
confirm Module flashing.
Note: Selecting No will result in further scanning and will not allow connection
of the rejected device to the Monitor.
Note: The green LED on the syringe and Module, respectively, will remain on
for 10 seconds when each device is connected to the Display.
Start Case
1) Enter the physician specified threshold. If no threshold is desired, enter "0".
ENSURE ALL PRIMING IS COMPLETED BEFORE PRESSING Start Case
ON THE Display .
For Detailed instructions, refer to the Contrast Monitoring Display or Smart
Monitor Instructions for Use.
2) Select Start Case on Display to begin contrast accounting.
3) Perform procedure
Pause/Resume
The DyeTect Pressure Module has a Pause button which allows the user to
manually pause/resume contrast accounting.
When the system is in "Pause" mode, cumulative contrast accounting is
suspended. "Resume" mode allows contrast accounting
1) To Pause select Pause on Display or press DyeTect Module pause button.
LED's on Smart Syringe & Module will blink yellow and background on
Display will change to yellow.
2) To Resume
Pause Lock Disabled
Aspirate contrast into Smart Syringe (resumes automatically) or select
Resume on Display or press Module pause button. LED's on Smart Syringe
and Module will be green for 10 seconds and the Monitor will return to normal
background.
Pause Lock Enabled
Select Resume on Display or press Module pause button. LED's on Smart
Syringe and Module will be green for 10 seconds and the Monitor will return to
normal background.
If contrast is injected back to the source, the system does NOT need to be
paused. The system will automatically recognize this as not being administered
to the patient. To maintain system accuracy, the user should ensure that only
100% contrast is injected back to the source.
Contrast Accounting Accuracy Tips:
Ensure system is Paused if contrast is not being injected to patient.
Ensure system is Active (unpaused) when contrast is being injected to
patient.
Ensure the minimum dwell time (1/2 second) is achieved between aspirations
of contrast and saline.
End Case
1) Press End Case button
2) When prompted, confirm that you want to end the case
Caution: Ending case will permanently disable communication between the
Monitor and disposables.
The image part with relationship ID rId18 was not found in the file.
Case summary screen appears showing cumulative volume of contrast
administered to patient and the percentage of physician specified threshold.
The Smart Syringe and Pressure Module contain batteries. Discard the Smart
Syringe and Pressure Module and single-use contrast source according to
hospital procedures.
Follow local governing ordinances regarding disposal. Do not incinerate as the
enclosed batteries may explode at excessive temperatures
Note: To view summaries for previous cases, from Main Menu select Settings,
User Settings, Case History.
Select Main Menu to return to the main menu and start a new case.
System Discontinuation
Smart Monitor: Power Down by pressing and holding the power button on the
end of the Monitor, then swiping the screen as indicated.