ERI or EOS Screens During MRI Mode

If MRI Mode has already been enabled and the Stimulator battery power falls below the threshold, the
Remote Control will display a message informing the patient that the Stimulator has entered the Elective
Replacement Indicator (ERI) period or has reached End of Service (EOS) of the device.
The patient can disable MRI Mode:
1. Press
to disable the MRI Mode.
2. Check the Remote Control to confirm that the Stimulator battery error message still
appears. See the "ERI (Elective Replacement Indicator) Screen (For Non-Rechargeable
Stimulators Only)" section of this manual.
Caution:
The Remote Control is MR Unsafe and must not be brought into the MRI scanner room.
Warning:
Do not perform an MRI scan if MRI Mode is not enabled. Scanning under different
conditions may result in patient injury or device malfunction.
ImageReady™ MRI Full Body Guidelines for WaveWriter Alpha™ and WaveWriter Alpha™ Prime Spinal Cord Stimulator Systems
MR System Preparation