1. Press
to dismiss the error message and return to the Remote Control Home Screen.
2. Instruct the patient to charge the Stimulator.
3. Enable MRI Mode once the Stimulator is fully charged.
Charge Stimulator Now Screen (Rechargeable Stimulators only)
If MRI Mode has already been enabled and the Stimulator battery power falls below the recommended
value, the Remote Control will display a message instructing the patient to charge the Stimulator.
To charge the Stimulator without disabling MRI Mode.
1. Do not press
2. Instruct the patient to charge the Stimulator.
3. Check the Remote Control to confirm that the error message has cleared.
4. Navigate to the Home Screen on the Remote Control by pressing
of the Remote Control and confirm that the MR Conditional Symbol
home screen.
The patient can also disable the MRI Mode before charging the Stimulator:
5. Press
to disable the MRI Mode.
6. Instruct the patient to fully charge the Stimulator.
7. Check the Remote Control to confirm that the error message has cleared.
8. Enable MRI Mode by following instructions in the Enabling MRI Mode section of this manual.
Caution: The Charger and Remote Control are MR Unsafe and must not be brought into the MRI
.
scanner room
ImageReady™ MRI Full Body Guidelines for WaveWriter Alpha™ and WaveWriter Alpha™ Prime Spinal Cord Stimulator Systems
MR System Preparation
button on the side panel