MR System Preparation
3. Select Yes
to proceed with enabling MRI Mode or No
to cancel the action.
Warning: An MRI scan is not recommended when the impedances are not within the acceptable range. Higher or lower than normal
impedances could indicate compromised Stimulator‑Lead integrity. Scanning under these conditions may increase the risk of
potential adverse effects listed in the "Safety Information" section under "Potential Interactions with MRI Environment."
Stimulator Error Screen
If the system check fails due to a Stimulator error, MRI Mode will not be enabled and the Remote Control
will display the Stimulator Error Screen. Do not perform an MRI scan if this error is displayed. Instruct the
patient to contact their physician managing their SCS System or Boston Scientific.
ImageReady™ MRI Full Body Guidelines for WaveWriter Alpha™ and WaveWriter Alpha™ Prime Spinal Cord Stimulator Systems
92395577-02
31 of 797