Component
Description
Surgical
IPG Port Plugs
Accessories
Clik Anchor
Clik X Anchor
Clik™ X MRI Anchor
Silicone Suture Sleeves
Med-A
Note: The system must be fully implanted and must include both an IPG and a Lead(s), at a minimum, to be MR Conditional.
The Lead(s) should be connected to the IPG, no Lead Extensions, Splitters, and Adapters are allowed. Leads implanted
without the IPG are not MR Conditional.
Note: For patients implanted with SC-2138-50, SC-2138-70 or SC-2208-50, SC-2208-70 Lead models:
• SC-2138-50 is the same Lead as SC-2158-50
• SC-2138-70 is the same Lead as SC-2158-70
• SC-2208-50 is the same Lead as SC-2218-50
• SC-2208-70 is the same Lead as SC-2218-70
ImageReady™ MRI Full Body Guidelines for WaveWriter Alpha™ and WaveWriter Alpha™ Prime Spinal Cord Stimulator Systems
92395577-02
4 of 797
Model Number(s) MRI System Settings
SC-4401
SC-4316
Surgical Accessories should
SC-4318
follow the MRI System Settings
SC-4319
used with the associated
N/A, included
implanted lead(s).
in kit
SC-4320